Causes of overheating of dental handpieces
High load use
Continuous use of the same handpiece for a long period of time (e.g., more than 10 minutes), the bearing produces mechanical fatigue due to high-speed rotation, resulting in increased friction and higher temperature.
Insufficient lubrication
Lack of timely lubrication of bearings increases friction and accelerates heating. Loss of lubricant or failure to replenish the oil in time after cleaning is a common cause.
Improper air pressure
Excessive air pressure (e.g., over 0.25 MPa for high-speed handpieces) accelerates bearing wear; low air pressure takes longer to complete the operation, indirectly leading to overheating.
Needle problems
The use of worn, bent or non-standard needles will increase cutting resistance and increase the load on the bearings.
Operating Practices: Measures to Avoid Overheating
Alternate use of multiple handpieces
Each handpiece should be used continuously for no more than 10 minutes. Alternating the use of handpieces reduces the mechanical fatigue of a single handpiece.
Control air pressure and speed
The working air pressure is recommended to be 0.20~0.25 MPa for high speed handpiece and 0.25~0.30 MPa for low speed handpiece, which needs to be adjusted according to the brand instruction.
Correct installation of the turning pin
Use standard needles (1.59~1.60 mm in diameter), avoid needles that are too long or too short.
Make sure the chuck is fully open when loading and unloading the needle to avoid damage to the clamping spring by idling.
Optimized operation
Use light touch and pointing operation, avoid sideways force or brutal operation to reduce bearing pressure.